Description
The AvCount2 determines the particle distribution in a liquid sample, whether checking the quality of fuel, filter systems or in-service lubricants, fast and precise results are produced. It’s compact design makes it ideal for laboratories with limited space; whilst its portable, rugged, stainless steel build also suits mobile testing in harsh environments.
- High visibility colour touch screen
- Portable design for laboratory and field use
- Rugged, stainless steel case
- Integrated printer
- 14 embedded test methods
Automatic test sequence
AvCount2 is fully automatic once the particle count test sequence is started. The instrument flushes the cell with sample prior to commencing the measurements. A precise volume of sample is then analysed. The flushing / sample analysing sequences are automatically repeated in accordance with the selected test method.
- The sample is drawn into analyser by an integral double pump
- Automatic changeover valve means that no operator intervention is required
- 10ml sample flows at 30ml/min through the measuring cell
- Particulate is measured by lightreduction across the cell
- Light reduction is proportional to particle size
- Results are automatically reported
- ASTM D7619 and IP565 results are automatically averaged
Results analysis
Once a test is completed, results are shown on screen for 6 particle size bands in Particles/ml and Cleanliness Code, these are stored within the AvCount2’s internal memory with space for up to 2000 sets of results. Results can be printed from the integral printer or downloaded to a USB memory stick to allow printing from a computer.